Temperature and Temperature/Humidity Test Chambers
The G-16 Elite
CE Certification on all Elite Models
|
Russells G-16 Elite Series temperature and temperature/humidity chambers represent one of our pre-engineered environmental chambers.
This model has a chamber volume in 16 cubic feet or 439 liters. Click for a complete overview highlighting the features and benefits
of our Elite Series as well as other standard model sizes to meet your simulation needs in testing extremes in cold and heat on your products. This reach-in chamber model is available with cryogenic cooling!
A Russells application engineer can assist you in selecting the laboratory chamber configuration, options, and size that's best for you,
custom designed to your exact specifications.
|
|
| |
|
| Chamber Model --- *denotes,
(B = temperature, D
= humidity) |
G*-16-3
|
G*-16-3-3
|
| Chamber Volume (cu ft) / (liters) |
16
/ 439 |
| Interior:
W x D x H |
(in) |
30 x 30
x 30 |
| (cm) |
76 x 76
x 76 |
| Exterior:
W x D x H |
(in) |
40 x 60
x 85 |
| (cm) |
102 x 153
x 216 |
| Shipping
Weight: |
(lb)
/ (kg) |
1,000
/ 450 |
1,200
/ 550 |
| Compressor(s): |
Hi
stage (hp) / lo stage (hp) |
3
/ N.A. |
3
/ 3 |
| Chamber
Heaters: |
(kw) |
6 |
6 |
| Humidity
Heaters: |
(GD
models only) (kw) |
2 |
2 |
| Power
Requirements: |
Voltage
/ Phase / Frequency |
230 / 3/
60 |
230 / 3
/ 60 |
| Load in (kva) |
15 |
19 |
| Performance: |
|
|
Temperature
Range
: |
˚F |
-22 to 350 |
-90 to 350 |
| ˚C |
-30 to 177 |
-68 to 177 |
| Pull-Down (chamber empty) |
In
Minutes |
| 75˚F
/ 23˚C |
to |
0˚F
/ -17˚C |
35 |
8 |
| to |
-40˚F
/ -40˚C |
N.A. |
12 |
| to |
-65˚F
/ -54˚C |
N.A. |
15 |
| to |
-85˚F
/ -65˚C |
N.A. |
20 |
| 350˚F
/ 177˚C |
to |
75˚F
/ 23˚C |
|
20 |
| Heat-Up (chamber empty) |
In
Minutes |
| 75˚F
/ 23˚C |
to |
240˚F
/ 116˚C |
18 |
18 |
| to |
350˚F
/ 177˚C |
25 |
25 |
| -90˚F
/ -68˚C |
to |
75˚F
/ 23˚C |
N.A. |
12 |
| Live load capacity (in watts) |
In
Watts |
| (steady
state) |
0˚F
/ -17˚C |
1,000 |
2,000 |
| -40˚F
/ -40˚C |
N.A. |
1,500 |
| -65˚F
/ -54˚C |
N.A. |
1,000 |
| Humidity
range (GD) Models: |
10% RH to 95% RH in the range bounded by a 185˚F
/ 85˚C maximum chamber temperature and a 40˚F / 4˚C
minimum dewpoint. |
|
|
|
|
|
|
|
 |
 |
  |
 |
| |
|